Animal weighing/averaging – Rice Lake Entris, Sartorius Laboratory User Manual

Page 27

Advertising
background image

Preparation

t

Select the Animal weighing application

in the menu:

see “Configuration"

t

Set the following parameters:

Applic.ation programs
Animalw. Animal weighing

Activty. Animal activity

calm Stable readout

o active Unstable

v.active Very unstable

Start

manual

o Autom.atic

o = Factory setting

Printout: Animal weighing

mDef 20

: Number of

subweighing
operations

x-Net + 410.1 g

: Calculated average

27

Animal Weighing/Averaging

Display symbol: V

Purpose

This application is used to determine the

weights of unstable samples (e.g., live

animals) or to determine weights under

very unstable ambient conditions. With

this program, the balance calculates the

weight as the average of a defined number

of individual weighing operations (also

referred to as “subweighing operations”).

Changing the Number of Subweighing

Operations

Activate function:

Press the b key

Select the desired number of measurements

(1 to 100):

In increments of 1: Press the b key

briefly

Increments of 10: Press and hold the
b

key.

The selected number of measurements is

stored in battery-backed memory.