Internal Calibration | Adjustment
Note:
Only for models with the label
Entris..i-1S !
Set the following parameters:
SETUP - BAL.SCAL.- CAL.Just. - Cal.Int. (Code 1.1. 9. 4)
The built-in motorized calibration weight located in the balance housing is applied and
removed automatically for internal calibration.
Step
Key (or instruction) Display
1. Tare the balance
w
0.0
g
2. Start calibration
Q
CAL.Int.
The internal calibration weight
is applied automatically.
CAL.RUN.
3. Adjustment carried out
CAL.end
4. The internal weight is
defined automatically
0.0
g