Rice Lake Entris, Sartorius Laboratory User Manual

Page 13

Advertising
background image

Internal Calibration | Adjustment

Note:

Only for models with the label

Entris..i-1S !

Set the following parameters:

SETUP - BAL.SCAL.- CAL.Just. - Cal.Int. (Code 1.1. 9. 4)

The built-in motorized calibration weight located in the balance housing is applied and

removed automatically for internal calibration.

Step

Key (or instruction) Display

1. Tare the balance

w

0.0

g

2. Start calibration

Q

CAL.Int.

The internal calibration weight

is applied automatically.

CAL.RUN.

3. Adjustment carried out

CAL.end

4. The internal weight is

defined automatically

0.0

g

13