14

IQ plus 590-DC Installation Manual

3.2.1

Configuration Menu

Figure 3-4. Configuration Menu

CONFIG Menu

Parameter

Choices

Description

Level 2 submenus

GRADS

10000
number

Graduations. Specifies the number of full scale graduations. The value entered must be in
the range 1–999 999 and should be consistent with legal requirements and environmental
limits on system resolution.

To calculate GRADS, use the formula, GRADS = Capacity / Display Divisions.

Display divisions for primary and secondary units are specified on the FORMAT menu.

ZTRKBN

OFF
0.5D
1D
3D

Zero track band. Automatically zeroes the scale when within the range specified, as long
as the input is within the ZRANGE and scale is at standstill. Selections are ± display
divisions. Maximum legal value varies depending on local regulations.

ZRANGE

1.9%
100%

Zero range. Selects the range within which the scale can be zeroed. The 1.9% selection is
± 1.9% around the calibrated zero point, for a total range of 3.8%. Indicator must be at
standstill to zero the scale. Use 1.9% for legal-for-trade applications.

Table 3-2. Configuration Menu Parameters

XXXXXXX

XXXXXXX

GRADS

10000

OFF

1D

0.5D

3D

ZTRKBN

ZRANGE

1.9%

100%

OFF

2D

1D

3D

MOTBAN

10D

5D

20D

FS+2%

FS+9D

FS+1D

FS

OVRLOA

1

2

4

8

DIGFL1

32

16

64

number

1

2

4

8

DIGFL2

32

16

64

1

2

4

8

DIGFL3

32

16

64

BOTH

PBTARE

NOTARE

KEYED

TAREFN

VERS

PROGRM

PFORMT

SERIAL

CALIBR

CONFIG

FORMAT

8OUT

32OUT

16OUT

64OUT

DFSENS

DFTHRH

2OUT

128OUT

4OUT

2DD

NONE

5DD

20DD

10DD

50DD

200DD

100DD

250DD

128

128

128

15HZ

3.75HZ

30HZ

7.5HZ

SMPRAT