3 revolution™ scaleware calibration, 3 revolution, Scaleware calibration – Rice Lake IQ plus 590-DC Installation Manual User Manual

Page 28

Advertising
background image

24

IQ plus 590-DC Installation Manual

2. Send the WZERO EDP command to calibrate

zero. The indicator displays

*CAL*

while

calibration is in progress.

3. Place test weights on the scale and use the

WVAL command to enter the test weight
value in the following format:

WVAL=nnnnnn<CR>

4. Send the WSPAN EDP command to calibrate

span. The indicator displays

*CAL*

while

calibration is in progress.

5. To remove an offset value, clear all weight

from the scale, including hooks or chains used
to hang test weights, then send the REZERO
EDP command. The indicator displays

*CAL*

while the zero and span calibrations are
adjusted.

6. Send the KUPARROW EDP command to exit

setup mode.

4.3

Revolution

Scaleware Calibration

To calibrate the indicator using Revolution Scaleware,
the indicator EDP port must be connected to a PC
running the Revolution configuration utility.

Use the following procedure to calibrate the indicator:

1. Place the indicator in setup mode (display

reads

CONFIG

) and remove all weight from

the scale platform.

2. Select

Calibrate Indicator

from the Revolution

Scaleware main menu.

3. On the Indicator Calibration display, select

t h e i n d i c a t o r m o d e l (

I Q + 5 9 0 - D C

) a n d

communications port, then click

OK

.

4. Revolution Scaleware uploads calibration

data from the indicator then presents the
information in a display like that shown in
Figure 4-2.

Figure 4-2. Revolution Scaleware Calibration Display

5. Enter the

Value of Test Weight

to be used for

span calibration then click

OK

.

6. The Zero Calibration dialog box prompts you

to remove all weight from the scale. Clear the
scale and click

OK

to begin zero calibration.

NOTE:

If your test weights require hooks or

chains, place the hooks or chains on the scale
for zero calibration.

7. When zero calibration is complete, the Span

Calibration dialog box prompts you to place
test weights on the scale for span calibration.
Place tests weights on the scale then click

OK

.

8. When span calibration is complete, the

Rezero dialog box prompts you to remove
weights from the scale. Remove the weights
then click

OK

.

9. When calibration is complete, the

New

Settings

fields of the Indicator Calibration

display are filled in. Click

Exit

to save the new

values and return to the Revolution Scaleware
main menu; to restore the previous calibration
values, click

Restore Settings

.