Rice Lake IQ plus 590-DC Installation Manual User Manual

Page 25

Advertising
background image

Configuration

21

REGULA

NONE
OIML
NTEP
CANADA

Regulatory mode. Specifies the regulatory agency having jurisdiction over the scale site.

• OIML, NTEP, and CANADA modes allow a tare to be acquired at any weight greater than

zero. NONE allows tares to be acquired at any weight value.

• OIML, NTEP, and CANADA modes allow a tare to be cleared only if the gross weight is at

no load. NONE allows tares to be cleared at any weight value.

• NTEP and OIML modes allow a new tare to be acquired even if a tare is already present.

In CANADA mode, the previous tare must be cleared before a new tare can be acquired.

• NONE, NTEP and CANADA modes allow the scale to be zeroed in either gross or net

mode as long as the current weight is within the specified ZRANGE. In OIML mode, the
scale must be in gross mode before it can be zeroed; pressing the ZERO key in net mode
clears the tare.

CONSNU

000000
number

Consecutive numbering. Allows sequential numbering for print operations. The consecutive
number value is incremented following each print operation.

The initial value of this parameter is set to the start up value specified on the CONSTU
parameter. Changing either CONSTU or CONSNU immediately resets the consecutive number
used for printing.

CONSTU

000000
number

Consecutive number start up value. Specifies the initial consecutive number (CONSNU) value
used when the indicator is powered on.

UID

1
number

Unit ID. Specify a unit identifier for the indicator in the range 1–999999. The unit ID can be
added to print ticket formats to identify the indicator used to generate the ticket.

Level 3 submenus

SBYDLY

0
number

Standby mode delay. Specifies the number of minutes the indicator must be inactive before
entering standby mode. Valid values are 0 (off) or 1–120 minutes.

The indicator enters standby mode if no key presses, serial communications, or scale motion
occur for the length of time specified on this parameter. Set this parameter to 0 to disable
standby mode.

LSTMOD

YES
NO

Last mode. Specifies whether the indicator resumes operation in the last mode it was in before
entering standby mode (YES) or performs the standard power-up sequence (NO).

ACCESS

DSABLE
ENABLE

Operator access to piece count mode. Specify DSABLE if piece count mode will not be used.
With access disabled, pressing the GROSS/NET (MODE) key toggles between gross and net
modes only.

SPLSIZ

10
20
50
100
5

Sample size. Specify the default sample size used for counting scale operations. Sample size
can be changed in counting mode during sample acquisition.

PROGRM Menu

Parameter

Choices

Description

Table 3-6. Program Menu Parameters (Continued)