4 serial menu – Rice Lake IQ plus 590-DC Installation Manual User Manual

Page 23

Advertising
background image

Configuration

19

3.2.4

Serial Menu

See Section 8.2 on page 34 for information about the IQ plus 590-DC serial data format.

Figure 3-7. Serial Menu

XXXXXXX

XXXXXXX

EOLDLY

000000

number

BITS

TERMIN

BAUD

CR

9600

7ODD

7EVEN

8NONE

CR/LF

4800

2400

1200

600

300

LFT

OFF

STREAM

INDUST

VERS

PROGRM

PFORMT

SERIAL

CALIBR

CONFIG

FORMAT

19200

SERIAL Menu

Parameter

Choices

Description

Level 2 Submenus

BAUD

9600
19200
300
600
1200
2400
4800

Baud rate. Selects the transmission speed for the EDP port.

BITS

8NONE
7EVEN
7ODD

Selects number of data bits and parity of data transmitted from the EDP port.

TERMIN

CR/LF
CR

Termination character. Selects termination character for data sent from the EDP port.

EOLDLY

0
number

End-of-line delay. Sets the delay period, in 0.1-second intervals, from when a formatted line is
terminated to the beginning of the next formatted serial output. Value specified must be in the
range 0-255, in tenths of a second (10 = 1 second).

STREAM

OFF
LFT
INDUST

Selects whether continuous data output is streamed from the serial port. Select LFT for
legal-for-trade applications in which the streamed data must match the current displayed weight.
Select INDUST to stream the latest weight data, regardless of the value displayed.

See Section 8.2 on page 34 for information about the IQ plus 590-DC continuous data format.

Table 3-5. Serial Menu Parameters