0 appendix, 1 error messages, 2 continuous output (stream) format – Rice Lake IQ plus 590-DC Installation Manual User Manual

Page 38

Advertising
background image

34

IQ plus 590-DC Installation Manual

8.0

Appendix

8.1

Error Messages

The IQ plus 590-DC provides a number of front panel error messages to assist in problem diagnosis. Table 8-1
lists these messages and their meanings.

8.2

Continuous Output (Stream) Format

Figure 8-1 shows the continuous output format sent to the IQ plus 590-DC serial port when the STREAM
parameter (SERIAL menu) is set to either LFT or INDUST.

Figure 8-1. Continuous Output Data Format

Error Message

Description

Solution

ADCERR

A/D not responding

Call Rice Lake Weighing Systems (RLWS) Service.

ADCKSM

A/D coefficient checksum error

CFCKSM

Configuration checksum error

Use TEST menu to perform DEFLT (restore defaults) procedure, then
recalibrate load cells.

CNTERR

Count error; insufficient sample size

Use larger parts sample.

DSPERR

Display error

Call RLWS Service.

EEPERR

EEPROM write error

LCCKSM

Load cell calibration checksum error

Recalibrate load cells.

LCPERR

Load cell power failed

Replace batteries. If error still shown, call RLWS Service.

OVERFL

Overflow error

Weight value too large (> 999999) to be displayed. Check
configuration.

PFCKSM

Print format checksum error

Check print formats for errors.

RAMERR

RAM test failed

Call RLWS Service.

TARERR

EEPROM read error

Use TEST menu to perform DEFLT (restore defaults) procedure, then
recalibrate load cells.

VIREE

EEPROM error

Use TEST menu to perform DEFLT (restore defaults) procedure, then
recalibrate load cells and reconfigure indicator.

------

Overrange error

Gross value exceeds overload limit. Check configuration.

______

A/D underrange

Weight value too small (< –99999) to be displayed or A/D reading <
–2.5 mV. Check load cell input signal.

STNDBY

Standby mode

Press any key to return to operating mode.

Table 8-1. IQ plus 590-DC Error Messages

<STX> <POL> <wwwwwww> <UNIT> <G/N> <S> <TERM>

ASCII 02

decimal

Polarity:
<space> = Positive
<–> = Negative

Weight: 7 digits, right-justified, dummy
zeroes, decimal point with no leading
zeroes except for leading zero immediately
preceding the decimal point. Leading
zeroes transmitted as spaces.

L = pounds
K = kilograms
T = tons
G = grains
<space> = grams
O = ounces

G = Gross

N = Net

Status:
<space> = valid
I = Invalid
M = Motion
O = Over/under range

<CR> <LF>

or <CR>

Advertising