23

4.0

Calibration

The IQ plus 590-DC can be calibrated using the front panel, EDP commands, or the Revolution

Scaleware

configuration utility. Each method consists of the following steps:

Zero calibration

Entering the test weight value

Span calibration

Optional rezero calibration for test weights using hooks or chains.

The following sections describe the calibration procedure for each of the calibration methods.

Figure 4-1. Calibration (CALIBR) Menu

4.1

Front Panel Calibration

To calibrate the indicator using the front panel, do the
following:

1. Place the indicator in setup mode (display

reads

CONFIG

) and remove all weight from

the scale platform. If your test weights require
hooks or chains, place the hooks or chains on
the scale for zero calibration.

2. Press until the display reads

CALIBR

(see

Figure 4-1). Press to go to zero calibration
(

WZERO

).

3. With

WZERO

displayed, press to calibrate

zero. The indicator displays

*CAL*

while

calibration is in progress. When complete, the
A / D c o u n t f o r t h e z e r o c a l i b r a t i o n i s
displayed. Press again to save the zero
calibration value and go to the next prompt
(

WVAL

).

4. With

WVAL

displayed, place test weights on

the scale and press to show the test weight
value. Use the numeric keypad to enter the
actual test weight, then press to save the
value and go to span calibration (

WSPAN

).

5. With

WSPAN

displayed, press to calibrate

span. The indicator displays

*CAL*

while

calibration is in progress. When complete, the
A / D c o u n t f o r t h e s p a n c a l i b r a t i o n i s
displayed. Press again to save the span
calibration value and go to the next prompt
(

REZERO

).

6. The rezero function is used to remove a

calibration offset when hooks or chains are
used to hang the test weights.

If no other apparatus was used to hang the
test weights during calibration, remove
the test weights and press to return to
the CALIBR menu.

If hooks or chains were used during
calibration, remove these and the test
weights from the scale. With all weight
removed, press to rezero the scale. This
f u n c t i o n a d j u s t s t h e z e r o a n d s p a n
calibration values. The indicator displays

* C A L *

w h i l e t h e z e r o a n d s p a n

calibrations are adjusted. When complete,
the adjusted A/D count for the zero
calibration is displayed. Press to enter
the value, then press to return to the
CALIBR menu.

7. Press until the display reads

CONFIG

, then

press to exit setup mode.

4.2

EDP Command Calibration

To calibrate the indicator using EDP commands, the
indicator EDP port must be connected to a terminal or
personal computer. See Section 2.3.3 on page 6 for
EDP port pin assignments; see Section 5.0 on page 25
for more information about using EDP commands.

Once the indicator is connected to the sending device,
do the following:

1. Place the indicator in setup mode (display

reads

CONFIG

) and remove all weight from

the scale platform. If your test weights require
hooks or chains, place the hooks or chains on
the scale for zero calibration.

XXXXXXX

XXXXXXX

WZERO

*CAL*

Display and edit

zero calibration

A/D count value

Display and edit

test weight value

WVAL

*CAL*

Display and edit

span calibration

A/D count value

WSPAN

VERS

PROGRM

PFORMT

SERIAL

CALIBR

CONFIG

FORMAT

*

CAL*

Press Enter to

remove offset from

zero and span

calibrations

REZERO