Introduction

1

About This Manual

This manual is intended for use by service technicians
responsible for installing and servicing IQ plus

®

590-DC digital weight indicators.

Configuration and calibration of the indicator can be
accomplished using the indicator front panel keys, the
EDP command set, or Version 2.2 or later of the
Revolution Scaleware

configuration utility. See

Section 3.1 on page 11 for information about
configuration methods.

Some procedures described in this
manual require work inside the indicator
enclosure. These procedures are to be
performed by qualified service
personnel only.

Authorized distributors and their employees
can view or download this manual from the
Rice Lake Weighing Systems distributor
site at

www.rlws.com

.

The

Operator Card

included with this manual

provides basic operating instructions for users of the
IQ plus 590-DC. Please leave the

Operator Card

with

the indicator when installation and configuration are
complete.

1.0

Introduction

The IQ plus 590-DC is a single-channel digital weight
indicator housed in a NEMA 4X/IP66-rated stainless
steel enclosure. The indicator front panel consists of a
1-inch (25 mm), six-digit, liquid crystal (LCD)
display and 19-button keypad. Features include:

Drives up to four 350

or eight 700

load

cells

Supports 4- and 6-wire load cell connections

Electronic data processing (EDP) port for full
duplex, RS-232 communications at up to
19200 bps

Powered by 6 C batteries for complete
portability

AC adapter for 115 or 230 VAC power

Configurable standby mode limits indicator
power consumption when scale is inactive

Piece count mode for basic parts counting
functions.

The IQ plus 590-DC is NTEP-certified for Classes III
and III L at 10,000 divisions. See Section 8.8 on
page 42 for detailed specifications.

1.1

Operating Modes

The IQ plus 590-DC has several modes of operation:

Normal (weighing) mode

Normal mode is the “production” mode of the
indicator. The indicator displays gross or net
weights as required, using the LCD annunciators
described in Section 1.3 on page 3 to indicate
scale status and the type of weight value
displayed. Once configuration is complete and a
legal seal is affixed to the back of the indicator,
the IQ plus 590-DC can operate only in normal
mode.

Piece count mode

In piece count mode, the indicator display shows
the number of parts on the scale rather than the
weight of those parts. Piece count mode has two
submodes:

Count display mode displays the current parts
count and allows ticket printing using the
CFMT print format.

Sample acquisition mode is used to calibrate
the indicator for parts counting.

Operator access to piece count mode is disabled
when the indicator is shipped from the factory.
See Section

6.0 on page

29 for detailed

information about counting operations.

Setup mode

Most of the procedures described in this manual
require the indicator to be in setup mode,
including configuration and calibration.

To enter setup mode, remove the large fillister
head screw from the enclosure backplate. Insert a
screwdriver or a similar tool into the access hole
and press the setup switch once. The indicator
display changes to show the word

CONFIG

.

Test mode

Test mode provides a number of diagnostic
functions for the IQ plus 590-DC indicator. Like
setup mode, test mode is entered using the setup
switch. See Section 8.7 on page 40 for more
information about entering and using test mode.

Standby mode

Standby mode is a configurable low-power mode
used to extend battery life when the indicator is
inactive. In standby mode, the LCD display shows
the word

STNDBY

; press any key to exit standby

mode. See Section

3.2.5 on page

20 for

information about configuring standby mode.