IQ plus 590-DC Installation Manual

5.1.2

Reporting Commands

Reporting commands (see Table 5-2) send specific
information to the EDP port. These commands can be
used in both setup mode and normal mode.

5.1.3

The RESETCONFIGURATION Command

The RESETCONFIGURATION command can be
used to restore all configuration parameters to their
default values. Before issuing this command, the
indicator must be placed in setup mode or test mode
(press and hold setup switch for approximately four
seconds to enter test mode).

Sending this command, followed by the RS (RESET)
EDP command, is equivalent to using the DEFLT
function on the TEST menu. See Section 8.7 on
page 40 for more information about test mode and
using the TEST menu.

NOTE:

All load cell calibration

settings are lost when the RESETCONFIGURATION
command is run.

5.1.4

Parameter Setting Commands

Parameter setting commands allow you to display or
change the current value for a particular configuration
parameter (Tables 5-3 through Table 5-8).

Current configuration parameter settings can be
displayed in either setup mode or normal mode using
the following syntax:

command<ENTER>

Most parameter values can be changed in setup mode
only. Use the following command syntax when
changing parameter values:

command=value<ENTER>

where

value

is either a number or a parameter value.

Use no spaces before or after the equal (=) sign. If you
type an incorrect command, the display reads

??

.

Changes to the parameters do not take effect until you
exit setup mode.

For example, to set the motion band parameter to 5,
type the following:

MOTBAND=5D<ENTER>

Command

Function

DUMPALL

List all parameter values

VERSION

Write IQ plus 590-DC software version

RS

Reset the indicator

P

Write current displayed weight with units
identifier.
The P command returns the current
displayed weight value to the EDP port,
along with the units identifier. If the indicator
is in an underrange or overload condition,
the weight value is replaced with ######
(overload) or %%%%%% (underrange).

S

Write one frame of stream format

Table 5-2. EDP Reporting Commands

Command

Description

Values

GRADS

Graduations

1–999 999

ZTRKBND

Zero track band

OFF, 0.5D, 1D, 3D

ZRANGE

Zero range

1.9%, 100%

MOTBAND

Motion band

1D, 2D, 3D, 5D, 10D, 20D, OFF

OVRLOAD

Overload

FS+2%, FS+1D, FS+9D, FS

SMPRAT

Sample rate

15HZ, 30HZ, 3.75HZ, 7.5HZ

DIGFLTR1
DIGFLTR2
DIGFLTR3

Digital filtering

1, 2, 4, 8, 16, 32, 64

DFSENS

Digital filter cutout sensitivity

2OUT, 4OUT, 8OUT, 16OUT, 32OUT, 64OUT, 128OUT

DFTHRH

Digital filter cutout threshold

NONE, 2DD, 5DD, 10DD, 20DD, 50DD, 100DD, 200DD, 250DD

TAREFN

Tare function

BOTH, NOTARE, PBTARE, KEYED

Table 5-3. CONFIG EDP Commands