3 calibration menu – Rice Lake IQ plus 590-DC Installation Manual User Manual

Page 22

Advertising
background image

18

IQ plus 590-DC Installation Manual

3.2.3

Calibration Menu

See Section 4.0 on page 23 for calibration procedures.

Figure 3-6. Calibration Menu

CALIBR Menu

Parameter

Choices

Description

Level 2 submenus

WZERO

Calibrate zero, display and edit the zero calibration A/D count value.

WVAL

Display and edit the test weight value.

WSPAN

Calibrate span, display and edit the span calibration A/D count value.

REZERO

Press Enter to remove an offset value from the zero and span calibrations.

Use this parameter only after WZERO and WSPAN have been set. See Section 4.1 on
page 23 for mor
e information about using this parameter.

Table 3-4. Calibration Menu Parameters

XXXXXXX

XXXXXXX

WZERO

*CAL*

Display and edit

zero calibration

A/D count value

Display and edit

test weight value

WVAL

*CAL*

Display and edit

span calibration

A/D count value

WSPAN

VERS

PROGRM

PFORMT

SERIAL

CALIBR

CONFIG

FORMAT

*

CAL*

Press Enter to

remove offset from

zero and span

calibrations

REZERO