0 operation, 1 power, 2 zero – Rice Lake MSI-8000HD Remote Display User Manual

Page 13: Operation, 1 power 2.2 zero

Advertising
background image

Operation 9

2.0

Operation

2.1 Power

To turn on the power.

1. Press

POWER

, the LCD will show all segments for a display text and the software

version number will display.

The MSI-8000HD is ready for use.

2.2 Zero

Sets the zero reading of the Scale/Dyna-Link. Takes out small deviations in zero when
the Scale/Dyna-Link is unloaded. (See Section 2.3 for zeroing (Taring) package,
rigging or pallet weights)

1. Press

ZERO

. The load reading must be stable within the motion window for

zero function to work.

The

MSI-8000HD

stores the zero reading, and can restore it even if power fails.

Zero - Rules for Use:

• Works in GROSS or NET mode.
• Zeroing while in Net mode will zero the gross load causing the display to show

the negative Tare value.

• The Scale/Dyna-Link must be stable within the motion window. They will not

zero until the stable annunciator is on. The Scale/Dyna-Link will “remember”
that it has a zero request for two seconds. If motion clears in that time, they will
zero.

• The Scale/Dyna-Link will accept a zero setting over the full Range of the Scale/

Dyna-Link. Zero settings above 4% of full Scale/Dyna-Link will subtract from
the overall capacity of the Scale/Dyna-Link.

Example
If 100 lb is zeroed on a 1000 lb Scale/Dyna-Link, the overall capacity of the
Scale/Dyna-Link will reduce to 900 lb plus the allowed over-range amount.