0 appendix, 1 troubleshooting, Appendix – Rice Lake MSI-8000HD Remote Display User Manual

Page 40

Advertising
background image

36 MSI-8000HD Operator’s Manual

6.0

Appendix

6.1 Troubleshooting

Problem

Possible Cause

Solution

Display is blank
when POWER
button is pressed

Discharged battery

Recharge

Defective battery

Replace (factory replacement only

Defective switch or
circuit board

Requires Authorized Service

Display does not
function properly/
Front panel
buttons do not
function normally/
Scale/Dyna-Link
will not turn off.

Improperly loaded
software

Reinstall software

Faulty Circuit Board

Requires Authorized Service

Loose Connectors

Requires Authorized Service

Scale/Dyna-Link
does not respond
to tension
changes

Out of calibration

Calibrate

Faulty load cell

Replace

Load cell connector

Check connector and wires

Display over
ranges below
100% of capacity

Tared Tension is
added to load to
determine overload
point

Return to gross tension mode

Zero requires
adjustment

Rezero the scale

Too much tension/
load has been zeroed

Rezero the scale

Display drifts

AZM (Auto0) is turned
off

Turn AZM on

Rapid temperature
changes such as
moving the scale from
indoors to outdoors

Wait until the scale temperature has
stabilized

Displayed tension
shows large error

Scale not zeroed
before load is lifted

Zero the scale with no load attached

lb/kg units causing
confusion

Select proper units

Requires recalibration

Recalibrate

Table 6-1. Troubleshooting