Qw  2/&qw, Access the service counters – Rice Lake MSI-8000HD Remote Display User Manual

Page 34

Advertising
background image

30 MSI-8000HD Operator’s Manual

1)

Program a User Function key to

be TEST (see function key setup).
F o r t h i s e x a m p l e ,

F 1

i s

programmed as TEST.

2)

Press

F1- TEST

.

3)

Within 2 seconds of pressing the

F1– TEST

key, press

F1

again

(must be F1 regardless of which key is

programmed as TEST).

The test will sequence through steps 4

to 7 automatically unless you stop it by

pressing

F2

immediately. Then using

F1

and

F2

each parameter can be

observed statically.

4)

The display flashes “LFCnt” (for

Lift Counter) followed by the
number of times the weight has
exceeded 25% of capacity.

5)

Next, the display flashes “OLCnt”

(for Overload Counter) followed
by the number of times the
weight has exceeded capacity.

6)

Next,the display flashes the C-

Cal value.

7)

The Dynalink / Scale returns to

standard weighing mode. If you
interrupted the auto sequence,
press

ZERO

to return to tension

link mode.

/)&QW 

2/&QW

Return to standard weight display



ENTER/SELECT

Number of Lifts above 25% of Capacity



Number of Lifts that exceeded Capacity

example value

example value

Start of Test Sequence

All segments on

&&D/

blinking

blinking

example C-Cal value

Access the Service Counters

Fx

Figure 5-8. Access the Service Counters

Only a MSI factory representative can reset the service counters, as these are an
important safety warning feature. Depending on the circumstances, a thorough load
train inspection is necessary to ensure product safety.
Reference MSI’s “Crane Scale Safety and Periodic Maintenance Manual” for proper
loading techniques to improve the safety and longevity of your Crane Scale or Dyna-
Link. This publication is available at www.msiscales.com and is included in the CD
shipped with your MSI-8000HD.