Vertical index error (v-index), Determining instrument errors, Checking and adjusting 55 – Leica Geosystems TPS110C User Manual

Page 55

Advertising
background image

Checking and Adjusting

55

TPS100z13

i

CALIBRATION 1/2

HZ-COLLIMATION

V-INDEX

[EXIT]

[OK]

Vertical Index Error (V-Index)

The vertical circle should read
exactly 90° (100 gon) when the
line of sight is horizontal. Any
deviation from this figure is
termed vertical index error (i).

By determining the vertical index
error the electronic level is
adjusted automatically.

Determining Instrument Errors

The calibration contains the
determination of the following
instrument errors:

Line-Of-Sight Error (Hz-
collimation)

Vertical Index Error (V-Index)
The electronic level is
adjusted simultaneously.

(

+

)

1. Open the menu
functions.
2. Select the menu
command
"CALIBRATION".

[OK] 3. Start "CALIBRATION"

[OK] 4. Start "HZ-

COLLIMATION"

or

4. Select the menu
command "V-INDEX".

[OK] 5. Start "V-INDEX"

Advertising