Determining the line-of-sight error (c), Determining the line-of-sight error (c) 57, Checking and adjusting 57 – Leica Geosystems TPS110C User Manual

Page 57

Advertising
background image

Checking and Adjusting

57

TPS100z17

ca. 100 m

± 4°30'

TPS100z18

180°

180°

HZ-COLLIMATION (1)

Hz

:

123°43'07"

V

:

272°11'31"

Sight target !

[EXIT]

[MEAS]

Determining The Line-Of-Sight Error (c)

1. Level up instrument exactly

using the electronic level.

2. Start Hz-collimation (refer to

p. 55).

3. Aim at a point approximately

100m from the instrument
that is less than ± 4°30' (5
gon) from the horizontal.

For checking the horizontal
aiming Hz- and V-angles are
displayed.

4. [MEAS] Trigger

measurement.

5. Change telescope face and

aim on point again.

6. [MEAS] Trigger

measurement again.

Advertising