Iii.3.2 the peltier temperature control device – Brookfield R/S CPS Rheometer User Manual

Page 12

Advertising
background image

Brookfield Engineering Labs., Inc.

Page 12

Manual No. M08-218-B0212

The following working fluids are usually used in the temperature bath:

-10°C to +90°C

Water (de-ionized) - Glycol Mixture

-20°C to +250°C

Silicone Bath Oil

Suitable bath fluids can be ordered from BROOKFIELD.

We recommend strongly that the upper temperature limit be set at

the liquid circulating thermostat to +95°C if water is used and to

+250°C if oil is used.

III.3.2 The Peltier Temperature Control Device

If the R/S CPS+ has been equipped with a Peltier temperature control system (i.e. model

R/S CPS+ P1 or R/S CPS+ P2), a bath/circulator is not used.

The Peltier temperature sensor Pt100 allows temperature regulation from 0°C to +135°C

(Model P1) or from +20°C to +180°C (Model P2).

A switching power supply (15V/60W) in the Peltier Unit P-TS supplies the Peltier ampli-

fier with power. A green light indicates that power is on while a red light indicates a fault.

For very low temperature operation with the P1 (under 10°C)

tap water circulating through the Peltier plate will improve

temperature control.

The Peltier Supporting Unit P-TS shows all existing control elements, indicators and con-

nections in Figures 4-A and 4-B.

Fig. 4-A: Front View of Peltier Supporting Unit P-TS