VIII. Technical Data

R/S-CPS+ Rheometer

Dimensions

480 mm x 300 mm x 290 mm

Weight

12 kg

Nominal operating voltage

Power consumption (average)

Power consumption (maximum)

± 15V, 5V

12W

22W

Ambient Conditions

Temperature

in operation

out of operation

Relative humidity (not condensable)

in operation

out of operation

10° to 40°C

10° to 45°C

20% to 80%

10% to 90%

Accuracy

± 3% of viscosity standard fluid value where M>30

± 1 digit

Torque range

1.5 to 50 mNm

Torque resolution

0.01 mNm

Speed range

0.01 to 1,000 min

-1

Angle resolution

15.7µrad

Temperature range

-20° to +250°C

depending on the geometry used

Range of shear rate

0 to 6,000 s

-1

depending on the geometry used

Range of shear stress

0 to 16,300 Pa

depending on the geometry used

Viscosity range

The given range is a standard value (not maximum value)

0.0008 to 125,000 Pa•s

depending on the geometry used.

Practical low limit is .050 Pa•s for cone/plate measurement.

AC Adapter

Dimensions

160 mm x 85 mm x 35 mm

Weight

0.5 kg

Power supply

Operating voltage

Output voltage

Output current

Output power

100 to 240 VAC

5V, ± 15V DC

2A, 0.9 / -0.2A

20W

Frequency

50 to 60 Hz

Ambient Conditions

Temperature

in operation

out of operation

Relative humidity (not condensable)

in operation

out of operation

+10° to +40°C

+10° to +45°C

20% to 80%

10% to 90%

Preset, Measured and Evaluated Values