VI.7.1 Configuration Output Mode

This function defines the output device (printer, serial interface) which will receive the re-

sults. This setting is independent of the internal memory in the rheometer for data storage.

When tests are run, the data is automatically stored in memory in addition to being sent to

the configured output device.
After selecting this function, you are prompted to select an output device:

Select output-device:
-No output-device
-Printer
-RS232

Internal memory only

Additional output

to serial interface

Additional output to

printer

Choice of an output device is stored in the instrument by pressing the OK key. This selec-

tion remains stored even after switching off the instrument!

The set output device is prompted before the start of every measurement (see Section VI.5.2:

Main Menu

Single measurement

and Section VI5.3: Main Menu

Run Program

).

Output of meas. val. to
-Printer
-Memory
<^>Menu <ST>Start

Output device now set

After selecting an output device and before the start of every measurement, be sure that:

a) the printer is connected to the rheometer and is ready for operation. If the printer is not

ready, you cannot start the measurement.

or

b) the RS232 is connected to the receiving device (normally a PC), which is set to the

data transfer parameters of the rheometer (see Section VI.7.5: Configuration RS232

Parameters

) and ready to receive data. If the receiving side is not ready, the data will

not be transmitted or will be transmitted to Null (see Section VI.8: Serial Data Transfer

via the RS232 Interface

).

VI.7.2 Configuration MeasCount Mode

The measurement series counter MEASCOUNT increases by 1 before each measurement

is started. The measurement series counter serves to identify a measurement series. The

counter increments until it is reset manually (“MeasCount=0”) or automatically. The user

defines whether the measuring counter is reset daily or not.