ERROR #1
Printer not ready!

<OK>cont. <ST>stop

If this error message appears, ensure the printer is ready and contains paper.

Press the

OK

key to try printing again or press the

ST

key to return to the menu.

VI.6.4 Utilities Measuring Systems

This function provides a method to create new measuring systems or change existing mea-

suring systems.
The following values can be edited:

• name of measuring system

• shear rate factor

K( [min/s]

(k_gamma)

• shear stress factor ‰ [Pa]

(tau_prom)

Only authorized personnel are permitted to change constants of the

measuring system!

After starting the function, you will be prompted to select the measuring system you want

to edit. Select the list item “NEW” to generate a new measuring system.

Select Meas. system:
1) C25-2/30

>

2) C25-1/30
3) C50-2/30

>

After selection of the measuring system entry, the user is prompted to enter a name, tau_prom

and k_gamma.

Enter Meas. syst. #4
Name: C50-1/30
tau_prom: 01.1418
k_gamma: 01.2910

The measuring system name is entered in alphanumeric form; the factors tau_ prom and

k_gamma are entered as numerical entries (see Section VI.4: Input of Numerical Values and

Alphnumeric Texts).
After entry, the system will prompt the user to store the new parameters.