Brookfield R/S CPS Rheometer User Manual

Page 35

Advertising
background image

Brookfield Engineering Labs., Inc.

Page 35

Manual No. M08-218-B0212

Available Measurement Types User Defined Range

Shear Rate D{s

-1

]

Depends on the measuring system

Speed n[rpm]

0.1 - 1,000 rpm

Shear Stress Tau [Pa]

Depends on the measuring system

Torque M[

]

0 - 999 promille (= 0 to 50 mNm)

Select the type of measurement and press the

OK

key (in our example “D[1/s]”). Now enter

the user defined values for each step one after the other.

The following inputs have to be made for each step:

Input Step #1
Start[1/s]:0010.00
End[1/s]: 0100.00
Nr of MP: 010

End Value

Start Value
Number of Measuring

Points

(data points)

Minimum and maximum start and end values depend on the selected measuring system for

shear rate (D[s

-1

]) and shear stress (Tau [Pa]). At input, the rheometer checks the start and

end values and indicates the message “Range error” if out of range:

Range error
Start value:
Min: 0.00
Max: 2400.00

Allowed minimum

value

Name of value out of

range

Allowed maximum

value

For example, a range error message would be displayed if a start value for D[s

-1

] is not

within the range of 0 - 2400.00 s

-1

when using measuring cone C50-2.

If the start value, end value and number of measuring points are acceptable, the user will

be prompted for Step Duration:

Input Step Nr. 1
Time[s]: 0060

Number of steps

being edited

Step duration

Minimum step duration:

Available Measurement Types User Defined Range

Shear Rate D{s

-1

]

tmin = number of measuring points * 4 s

Speed n[rpm]

tmin = number of measuring points * 4 s

Shear Stress Tau [Pa]

tmin = number of measuring points * 1 s

Torque M[

]

tmin = number of measuring points * 1 s