VII. Measuring

VII.1 Measuring in Manual Operation

You can measure in manual mode by following these brief instructions:

• Install the R/S-CPS+ Rheometer (see Section III: Instrument Installation).

• Connect the AC adapter (see Section III.2.2: AC Adapter).

• Connect the printer, which is optional (see Section III.2.3: Printer Connection).

• Fill and mount the measuring system (see Section V: Measuring Systems).

• Make sure no substance or solvent enters the measuring element coupling, the measuring

drive or the electronic unit.

• Wait until the sample reaches the desired temperature.

• Start a Single Point Test or Run Program (see Section VI.5: Menu Entries).

• After running the test, remove the sample and clean the measuring system.

VII.2 Measuring in Remote Operation

• Install the R/S-CPS+ Rheometer (see Section III: Instrument Installation).

• Connect the AC adaptor (see Section III.2.2: AC Adapter).

• Connect the RS232 cable between the rheometer and the PC.

• Switch on the R/S-CPS+ Rheometer and select the menu level “Remote” on the main screen.

• Switch on the computer and all peripheral devices.

• Start the Rheo 3000 software.

• Load a program within the Rheo 3000 software.

• Fill and mount the measuring system (see Section V).

• Make sure no substance or solvent enters the measuring element coupling, the measuring

drive or the electronic unit.

• Wait until the sample reaches the desired temperature.

• Start a Single Point Test or Run Program (see Section VI.5: Menu Entries).

• After running the test, remove the sample and clean the measuring system.