Brookfield R/S CPS Rheometer User Manual

Page 50

Advertising
background image

Brookfield Engineering Labs., Inc.

Page 50

Manual No. M08-218-B0212

Cause:

An impermissibly high value was measured during the zero point cali-

bration of the rheometer.

Example

Message

Error #3
Zero cal. error!
Please retry cal.
<OK> cont. <ST> stop

What to Do: a] press the

ST

key

b] ensure the measuring system has been removed from the rheometer

c] retry zero point calibration - if error occurs again, this is an indication

of a fault in the measuring instrument

Cause:

The maximum torque has been exceeded during a measurement. Pos-

sible reasons:

a] the measuring range is inappropriate for the measured sample

b] the cone or plate spindle cannot rotate on the measuring plate

Example

Message

ABORT:torque max!
Single measurement
total time: 60 s
total no. of MP: 10

What to Do: a] press the

OK

key

b] check whether the measuring system has been fastened properly at the

rheometer

c] if fastened properly, select smaller shear rates or speeds for this mea-

surement or use a measuring system with a higher shear stress factor

ABORT: Torque Max

Zero Calibration Error