A.2 Error Messages

The R/S-CPS+ Rheometer is user-friendly regarding errors. When errors occur, they are

trapped, and the user is informed on the LCD. The most frequent error messages are explained

in this chapter.

Cause:

User’s error at input. The user has tried to input a value that is less than

the allowed minimum value or is more than the allowed maximum value.

Example of

Faulty Input

of a Preset

Value:

Range error
End value:
Min: 0.90
Max: 1032.80

Allowed

minimum value

Name of value out

of range

Allowed maximum

value

What to Do: Enter the value again. Be sure that the new value is within the allowed

value range.

Cause:

The printer has been selected as the output device but is not ready for

printing. Possible reasons include:

a] printer cable is not connected to the rheometer

b] paper is out, printer is not on line, other printer errors

Example of

Print Error

Message

Error #1
Printer not ready!

<OK> cont. <ST> stop

Try again

Abort activity;

do not try again

What to Do: a] make sure printer cable is connected to the rheometer

b] make sure paper tray has paper

c] check that the “ready” LED lights of the printer are on

d] remove the error cause at the printer and press the

OK

key at the rhe-

ometer

e] press the

ST

key to stop the process if you cannot find the error cause.

Range Error

Printer Error

Advertising