Vi.5.2 main menu → single measurement – Brookfield R/S CPS Rheometer User Manual

Page 26

Advertising
background image

Brookfield Engineering Labs., Inc.

Page 26

Manual No. M08-218-B0212

THERMOSTAT

TS: 20.0
T: 20.0
Status: 00

Permanent Indication of

Actual Temperature Value

Adjustable Temperature

Set Value

Error Indication

(00 = no error)

p

/

q

Info on Temperature

Value That Can Be Varied

The last adjusted set point TS is always saved when the rheometer is turned off. When the

rheometer is re-started, the Peltier will return to the last set point.

Modify the set point temperature with the and keys.
The temporary flashing of three points “• • •” confirms the acceptance of the new set tem-

perature value:

THERMOSTAT

TS: 45.0 ...
T: 20.0
Status: 00

p

/

q

Confirmation of Modified

Set Temperature Value

Adjustment of the actual temperature value starts automatically. The main menu can be
re-entered with the

ST

and

OK

keys.

VI.5.2 MAIN Menu

Single Measurement

This function measures shear stress or shear rate (CSS or CSR) at constant user defined

values. The user may select a choice of measurement method from the follwoing:

Shear Rate

D [s-1]

Speed

n [U/min] or [rpm]

Shear Stress

Tau [Pa}

Torque

M[

] [1000

correspond to 50 mNm]

Before you can start the measurement itself, you have to input the following values:

• Selection of the measuring system used (see Section VI.3: Selecting from Lists).

• Selection of test method (see Section VI.3: Selecting from Lists). This is where you de-

cide if you want a controlled rate measurement (shear rate, RPM) or a controlled stress

measurement (shear stress, torque).

• Input of pre-set values (see Section VI.4: Input of Numerical Values and Alphanumer-

ic Texts

).

- D [s

-1

]

range of values depends on the measuring system

- n [rpm]

0.7 to 1,000 rpm

- Tau [Pa]

range of values depends on the measuring system

- M[‰]

0-999 ‰

• Input of desired number of measuring points (see Section VI.4: Input of Numerical