Measurements can be taken with the memory full but the results will only be shown at the

LCD and will be lost after completion of the measurement. If you try to send results to a

printer and it is not connected or has no paper in it, an error message will be displayed until

the printer is connected and operational or until you stop the activity.

When you start the measurement, the instrument shows:

Program running...
Wait for 1.MP

#01

This message will be displayed until the first measuring point is displayed.

100.5

010.00 1/s

Eta: 1.728 Pas
Tau: 1.572 Pa
05s 20.1°C

#01

Torque

Duration

Temperature

Actual pre-set value

Measurement result

Current step

If the torque is less than 30‰, the results may not have the desired accuracy or repeatability.

The user should consider changing the user defined test parameters to get a torque higher

than 30‰.
If the display field for the torque indicates: “Mlow!”, these values are below the range of

resolution of the rheometer.
If the temperature is not to displayed, the measuring sensor is not connected and the printout

of the temperature will be printed as “---”.
The display is updated with every new measuring point. The current measurement can be

cancelled at any time with the

ST

key.

After measurement, or after a stop, the display field for step indicates “END” or “BREAK”.

The LCD alternates at intervals of about four seconds between the last displayed measuring

point and information on the measurement:

Program end
Single measurement
Total time: 200s
Total #MP: 40

Reason for

break or end

Duration until

break or end

Type/name of

executed program

Number of measurement

points

Pressing the

OK

key stops the alternating display and returns the user to the main menu.