Vi.6 menu entries in the utilities menu, Vi.6.1 utilities → zero calibration – Brookfield R/S CPS Rheometer User Manual

Page 32

Advertising
background image

Brookfield Engineering Labs., Inc.

Page 32

Manual No. M08-218-B0212

VI.6 Menu Entries in the Utilities Menu

The Utilities menu contains several useful functions.

Zero Calibration

Initiates the zero calibration procedure of the rheometer

Edit Programs

Input or modification of programs which are started with Run Pro-

gram

Print Programs

Prints all parameters of all the programs in memory to the printer

Measuring Systems Input of the measuring system parameters or generation of new

measuring systems

Print Memory

Output of data in memory to the printer or to the serial interface

Delete Memory

Clears all data from memory

Temperature Meas. Measures temperature without running a measurement

The following sections explain these options in more detail.

VI.6.1 Utilities Zero Calibration

The Zero Calibration function serves to calibrate the zero point of the rheometer. This

function continues for approximately 10 minutes and should be done once a week.

Before starting this function, be sure the R/S-CPS+ Rheometer has

warmed up for at least 10 minutes, and that NO measuring element is

in the measuring element coupling.

To confirm the command that the measuring system is removed, press the OK key (start of

zero point calibration), or press the ST key to return to the Utilities menu.
Zero point calibration will proceed automatically and comprises several measuring series at

different speeds. The progress of calibration is shown by the number of executed steps of

the total steps. If an error message appears during zero calibration, repeat the zero calibra-

tion. If the error message is displayed again there may be a technical fault (See Appendix

G: Warranty and Repair Services).

After successful calibration, the values of the zero point are stored internally. These values

are preserved until the next calibration.

Press the OK key to save the values. Press any other key to return to the Utilities menu

without saving.

Note: You can cancel zero point calibration at any time with the ST key. The zero point

values determined up to that point will be ignored.