Mmenu, Setup, Reset – Holtgreven Ohaus Defender Series Bench Scales User Manual

Page 22: D.scale, Range1

Advertising
background image

EN-20 7000 Series Indicators

CAL

GEO

CAL

END CAL

GEO

12

mMENU

SETUP

SEtUP

RESET

YES

SEtUP

DUAL SCALE

rESEt

NO

ON

SEtUP

RANGE1

DUAL

3.3.6 Geographical Adjustment Factor
Geographical Adjustment Factor (GEO) is used to adjust the calibration based on the current
location. Refer to Section 6, table 6-1 and set the GEO factor that corresponds to your location.

Settings from 1 to 31 are available.

Press the

No or Back button to change the value. Press the Yes button to accept the value.


NOTE: Only an authorized manufacturer’s representative or certified verification personnel may make
these changes. Changing the geographical setting alters the calibration values.


3.3.7 End Calibration
Press the

Yes button to advance to the next menu or the No button to return to the top of the current

menu.

3.4

Setup Menu

When the indicator is used for the first time, enter this menu to set the Range, Capacity and
Graduation values. If the indicator comes as part of a bench scale, these values were already set in
the factory.


NOTE: If two scales will be operated, set DUAL SCALE to ON and set the Range, Capacity and
Graduation values for both scales. The Range1, Capacity1 and Graduation1 values are set for
Scale 1, which is connected to the circular connector (Figure 1-1, Item 8) or terminal block J4
(Figure 1-3, Item 9) on the main pc board. The Range2, Capacity2, and Graduation2 values are
set for Scale 2, which is connected to the terminal block J7 (Figure 1- 3, Item 5) on the main pc

board. All other Setup Menu settings apply to both scales.

3.4.1 Reset
Reset the Setup menu to the factory defaults. Factory default settings are shown in bold.

NO

= not reset

YES

=

reset






3.4.2 Dual Scale
Set the status of the second scale input (Scale 2).

OFF

= Scale 2 is not active.

ON

= Scale 2 is active.


NOTES: If DUAL SCALE is set to OFF, the RANGE, CAPACITY and GRADUATION menu items are

displayed.
If DUAL SCALE is set to ON, the RANGE1, RANGE2, CAPACITY1, CAPACITY2, GRADUATION1
and GRADUATION2 menu items are displayed, where the number represents the scale
number.


3.4.3 Range1
Set the number of weighing ranges for Scale 1.

SINGLE

= one weighing range from zero to capacity.

DUAL

= two weighing ranges. The fine range (1r) is from zero to half
capacity. The coarse range (r2) is from half capacity to full capacity.

NOTE: if DUAL SCALE is set to OFF, RANGE is displayed instead of RANGE1.

d.SCALE

OFF

rANGE1

SINGLE

Advertising