Setup, Cap2, Range2 – Holtgreven Ohaus Defender Series Bench Scales User Manual

Page 23: Cap1, Grad1

Advertising
background image

7000 Series Indicators EN-21

SEtUP

RANGE2

DUAL

SEtUP

CAPACITY1

999999

SEtUP

CAPACITY2

CAP2

kg

999999

100

SEtUP

GRADUATION1

999999

999999

3.4.4 Range2
Set the number of weighing ranges for Scale 2.

SINGLE

= one weighing range from zero to capacity.

DUAL

= two weighing ranges. The fine range (1r) is from zero to half

capacity.

The coarse range (r2) is from half capacity to full capacity.



3.4.5 Capacity1
Set the capacity of Scale 1 using the numeric keypad.

Settings from

1 to 999999 are available.


NOTE: if DUAL SCALE is set to OFF, CAPACITY is displayed instead of CAPACITY1.

NOTE: If DUAL was selected as the Range1 setting, the Capacity1 setting defines the capacity of the
coarse range. The capacity of the fine range is automatically defined as half of the Capacity1
setting. For example, if the Capacity1 setting is 15, the capacity of the fine range is 7.5 and the

capacity of the coarse range is 15.

After the capacity is set, select the primary unit.

kg

= the primary unit is kilograms.

lb

= the primary unit is pounds.




3.4.6 Capacity2
Set the capacity of Scale 2 using the numeric keypad.

Settings from

1 to 999999 are available.


NOTE: If DUAL was selected as the Range2 setting, the Capacity2 setting defines the capacity of the
coarse range. The capacity of the fine range is automatically defined as half of the Capacity2
setting.

After the capacity is set, select the primary unit.

KILOGRAM

= the primary unit is kilograms.

POUND

= the primary unit is pounds.




3.4.7 Graduation1
Set the readability of Scale 1.

Settings of 0.0001, 0.0002, 0.0005,

0.001, 0.002, 0.005, 0.01, 0.02, 0.05, 0.1, 0.2,

0.5, 1, 2, 5, 10, 20, 50, and 100 are available.


NOTE: if DUAL SCALE is set to OFF, GRADUATION is displayed instead of GRADUATION1.

NOTE: Graduation1 setting selections are dependent on the Capacity1 setting. Selections are limited
to values that will provide a scale resolution between 1:1000 and 1:50000.


NOTE: If DUAL was selected as the Range1 setting, the Graduation1 setting defines the fine range
graduation. The coarse range graduation is automatically defined as one step greater than the
Graduation1 setting. For example, if Graduation1 is set to 0.001, the coarse range graduation is
defined as 0.002.

rANGE2

SINGLE

CAP1

kg

1

CAP1

lb

1

CAP2

kg

1

CAP2

lb

1

Grad1

kg

0.0001