EN-64 7000 Series Indicators

8.

TECHNICAL DATA

8.1

Specifications

The technical data is valid under the following ambient conditions:
Temperature: -10°C to 40°C / 14°F to104°F
Relative humidity: Maximum relative humidity 80% for temperatures up to 31°C decreasing linearly to 50% relative humidity at
40°C.
Altitude: up to 2000m

TABLE 8-1. SPECIFICATIONS

Indicator Model

T71P

T71XW

Maximum displayed resolution

1:50,000

Maximum approved resolution

1:10,000

Maximum counting resolution

1:500,000

Weighing units

Kilogram, Gram, Pound, Ounce, Pound:Ounce, Tonne, Custom

Weighing modes

Weighing, Parts Counting, Percent Weighing, Check Weighing,

Dynamic Weighing / Display Hold

Features

Accumulation statistics, Library record storage, Under / Accept / Over LEDs

Display

25 mm high, 2-line LCD

Under/Accept/Over indicators

Yellow, Green, Red LED

Backlight White

LED

Controls

17 button membrane switch

Ingress protection

---

IP66

Load cell excitation voltage

5 VDC

Load cell drive

Up to 8 x 350 ohm load cells

Load cell input sensitivity

Up to 3 mV/V

Stabilization time

Within 2 seconds

Auto zero tracking

Off, 0.5 d, 1 d or 3 d

Zeroing range

2% or 100% of capacity

Span calibration

1 kg or 1 lb to capacity

Housing dimensions (W x D x H)

260 x 71 x 168 mm

10.2 x 2.7 x 6.6 in

262 x 76 x 149 mm

10.3 x 3.0 x 5.8 in

Net weight

1.5 kg

3.3 lb

3.5 kg

7.7 lb

Shipping weight

2.3 kg

5 lb

4.3 kg

9.5 lb

Operating temperature range

-10 °C to 40 °C

14 °F to 104 °F

Mains power

100-240 VAC / 50-60 Hz internal power supply

Overvoltage category

II

Pollution degree

2

Battery power

6 C-size (LR14) batteries (not supplied)

Rechargeable battery pack (option)

Rechargeable battery pack (option)

Interfaces RS232

(included)

External Input (included)

Second RS232 (option)

RS485/RS422 (option)

Advertising