Setup, A.tare, Grad2 – Holtgreven Ohaus Defender Series Bench Scales User Manual

Page 24: Pwwr.un, Zero, Delay

Advertising
background image

EN-22 7000 Series Indicators

SEtUP

GRADUATION2

100

SEtUP

POWER UNIT

SEtUP

ZERO RANGE

SEtUP

AUTO TARE

A.tArE

ACCEPT

100

CUSTOM

TONNE

POUND-OUNCE

OUNCE

POUND

GRAM

KILOGRAM

ACCEPT

ON

5

2

1

0.5

3.4.8 Graduation2
Set the readability of Scale 2.

Settings of 0.0001, 0.0002, 0.0005, 0.001, 0.002, 0.005, 0.01, 0.02, 0.05, 0.1, 0.2,

0.5, 1, 2, 5, 10, 20, 50, and 100 are available.


NOTE: Graduation2 setting selections are dependent on the Capacity2 setting. Selections are limited
to values that will provide a scale resolution between 1:1000 and 1:50000.

NOTE: If DUAL was selected as the Range1 setting, the Graduation2 setting defines the fine range
graduation. The coarse range graduation is automatically defined as one step greater than the
Graduation2 setting.

3.4.9 Power Unit
Set the unit of measure displayed at startup.

AUTO

= last unit in use when the indicator was turned off.

KILOGRAM =

kilograms

GRAM

=

grams

POUND

= pounds

OUNCE

= ounces

POUND OUNCE = pound ounces
TONNE

= tonnes

CUSTOM

= custom unit





3.4.10 Zero Range
Set the percentage of scale capacity that may be zeroed.

2%

= zero up to 2 percent of capacity.

100%

= zero up to 100 percent of capacity.




3.4.11 Auto Tare
Set the automatic tare functionality.

OFF

= automatic tare is disabled.

ON

= the first stable gross weight is tared.

ACCEPT

= in Check weighing mode, stable gross loads within the accept limits

are tared.




When ACCEPT is selected, set the delay time.

OFF

= automatic tare takes affect as soon as the weight is stable.

0.5

= automatic tare takes affect after the weight is stable for 0.5 second.

1

= automatic tare takes affect after the weight is stable for 1 second.

2

= automatic tare takes affect after the weight is stable for 2 seconds.

5

= automatic tare takes affect after the weight is stable for 5 seconds.



Grad2

kg

0.0001

PwWr.UN

AUTO

ZErO

%

2

A.tArE

OFF

dELAY

OFF