Diff, Contnt, Info – Holtgreven Ohaus Defender Series Bench Scales User Manual

Page 36: Weighing none, Counting, Percent, Check weighing, Dynamic, Difference, Information

Advertising
background image

EN-34 7000 Series Indicators

ON

ON

Difference
Set the status.

OFF

= the difference is not printed following the Calibration Test procedure.

ON

= the difference is printed following the Calibration Test procedure.


--------- Cal Test ---------
New Cal: 10.000 kg
Old Cal: 9.999 kg
Diff: 0.001 kg
Wt. ID:------------------
------------ End -----------


Information
Set the status.

OFF

= the reference information is not printed.

ON

= the reference information is printed.


NOTE: The reference information is dependent on the mode and the ACCUMULATE setting. Examples
are shown below.

Mode

Accumulate set OFF

Accumulate set ON

Weighing

None


N: 10
Total: 10.000 kg
Avg: 1.000 kg
Std: 0.001 kg
Min: 0.999 kg
Max: 1.001 kg
Diff: 0.002 kg

Counting


APW: 0.100 kg


APW: 0.100 kg
N: 10
Total: 1000 Pcs
Avg: 100 Pcs
Std: 1 Pcs
Min: 99 Pcs
Max: 101 Pcs
Diff: 2 Pcs

Percent


Ref Wt: 1.23 kg


Ref Wt: 1.23 kg

Check Weighing


Under: 0.995 kg
Over: 1.005 kg


Under: 0.995 kg
Over: 1.005 kg
N: 10
Total: 10.000 kg
Avg: 1.000 kg
Std: 0.001 kg
Min: 0.999 kg
Max: 1.001 kg
Diff: 0.002 kg

Dynamic


Level: 0


Level: 0
N: 10
Total: 10.000 kg
Avg: 1.000 kg
Std: 0.001 kg
Min: 0.999 kg
Max: 1.001 kg
Diff: 0.002 kg

dIFF

OFF

CONtNt

DIFFERENCE

INFO

OFF

CONtNt

INFORMATION

Advertising