EN-60 7000 Series Indicators

mMENU

AUdIt

CFG000

CAL000

*

0

0.000

kg

*

0

0.000

kg

When the scale base is attached to the indicator using a connector, it is necessary to seal the load cell cable to the indicator in
some jurisdictions. The load cell sealing collar P/N 80500737 (Figure 6-5) is available as an accessory.

Figure 6-5. T71P Load Cell Sealing Collar

6.3.2 Audit Trail Seal

For jurisdictions that use the audit trail sealing method, the local weights and measures official or authorized service agent must
record the current configuration and calibration event counter values at the time of sealing. These values will be compared to
values found during a future inspection.

NOTE: A change to an event counter value is equivalent to breaking a physical seal.

The audit trail uses two event counters to record changes to configuration and calibration settings.
• The configuration event counter (CFG) will index by 1 when exiting the menu if one or more of the following settings are

changed: Range, Capacity, Graduation, Power On Unit, Zero Range, Auto Tare, Legal for Trade, Stable range, Auto Zero
Tracking, Modes, Units, Stable Printing Only. Note that the counter only indexes once, even if several settings are changed.

The configuration event counter values range from CFG000 to CFG999. When the value reaches CFG999, the count starts
over at CFG000.

• The calibration event counter (CAL) will index by 1 when exiting the menu if a Span Calibration, Linearity Calibration or GEO

setting change is made. Note that the counter only indexes once, even if several settings are changed. The calibration event

counter values range from CAL000 to CAL999. When the value reaches CAL999, the count starts over at CAL000.


The event counters can be viewed by pressing and holding the MENU button.


While the button is held, the display will show MENU followed by Audit.



Release the button when Audit is displayed to view the audit trail information.



The audit trail information is displayed in the format CFGxxx and CALxxx.





Then the indicator returns to normal operation.