Holtgreven Ohaus Defender Series Bench Scales User Manual

Page 58

Advertising
background image

EN-56 7000 Series Indicators

5.

SERIAL COMMUNICATION

5.1

Interface Commands

The indicator can be controlled using the commands listed below.

Command
Characters

1)

Function

ON

Turns the indicator on.

OFF

Turns the indicator off.

IP

Immediate print of displayed weight (stable or unstable).

P

2)

Print displayed weight (stable or unstable).

SP

Print on stability.

CP Continuous

print.

xP

Print on interval, where x = 1 to 3600 (seconds).

Z

2)

Equivalent to pressing the

ZERO button.

T

2)

Equivalent to pressing the

TARE button.

xT

Establish a preset tare, where x = the tare value in the current weighing unit.

PU

Print the current weighing unit.

xU

Change the weighing unit, where x = 1 (g), 2 (kg), 3 (lb), 4 (oz), 5 (lb:oz), 6 (t), 7 ©.

PV

Print the name, software version and LFT ON (if LFT is set to ON).

H x “text”

Enter the header line, where x = 1 to 5 (line number) and “text” = header text up to 24 characters.

Escape key and R Global reset (all menu settings are reset to their factory default settings).

NOTES:

1) Commands sent to the indicator must be terminated by a carriage return (CR) or a carriage return-line feed (CRLF).
2) Alternate command characters may be defined by the user. Refer to Section 3.10.7.
3) Data output by the indicator is always terminated with a carriage return-line feed (CRLF).

5.2

Output Format

The Result Data is output in the following format.

Field: Label

1

Space

2

Weight

3

Space

2

Unit

4

Space Stability

5

Space G/N

6

Space Term. Characters

7

Length: ≤11

≤1

9

≤1

≤ 5 1

≤ 1

≤ 1

≤ 3 ≤ 1

≤ 4

1) In certain cases, a Label field of up to 11 characters is included. Refer to Section 5.3.

2) Each field is followed by a single delimiting space (ASCII 32).
3) The Weight field is 9 right justified characters. If the value is negative, the “-“ character is located at the immediate

left of the most significant digit.

4) The Unit field contains the unit of measure abbreviation up to 5 characters.
5) The Stability field contains the “?” character if the weight reading is not stable. The Stability field and the following

Space field are omitted if the weight reading is stable.

6) The G/N field contains the net or gross indication. For net weights, the field contains “NET”. For gross weights, the

field contains nothing, “G” or “B”, depending on the GROSS INDICATOR menu setting. Refer to Section 3.5.8.

7) The Termination Characters field contains CRLF, Four CRLF or Form Feed (ASCII 12), depending on the LINE FEED

menu setting. Refer to Section 3.9.5.