1 setting the heating profile – KERN & SOHN DLT 100-3N User Manual

Page 41

Advertising
background image

DLT_N-BA-e-1421

41

11.1.1 Setting the heating profile

This function offers different heating profiles to adapt the drying characteristic
optimally to the used sample.

Tap

< Heating profile >

in dropdown menu

(see chap. 11.1).

The available heating profiles are displayed:

Standard

Fast

Ramp

Steps

1. Standard drying

Display pictogram during drying

This heating profile are used for most
samples. The sample will be heated up to the
set drying temperature,
selectable 35°C -160°C

Tap < Standard >.

The numeric input window will be displayed.
Enter required drying temperature,
selectable 35°C – 160°C.

Import entry with

<OK>

, the display returns to

the previous screen.

2. Quick drying

Display pictogram during drying

This heating profile is suitable for samples
with high moisture content (e.g. liquids).
After start the selected temperature will be
exceeded for a short time - to compensate
the evaporation coldness and to accelerate
the drying process.
Then the temperature is automatically
controlled down to the set value.

Tap < Fast >.

The numeric input window will be displayed.
Enter required drying temperature, selectable
35°C – 160°C.

Import entry with

<OK>

, the display returns to

the previous screen.

Advertising