3 preheating (before weighing-in) – KERN & SOHN DLT 100-3N User Manual

Page 46

Advertising
background image

46

DLT_N-BA-e-1421

11.1.3 Preheating (before weighing-in)

If needed, the function „preheating“ can be enabled for preheating the sample
chamber prior to beginning the actual measurement.

A preheated device may affect the consistency of the results as the device keeps the
same temperature status for each measurement. Preheating will take place up to the
point of measuring, in standby mode (For setting see chap. 7, [2]) including after
finished measuring.

Tap < Preheating > in dropdown menu
(See chap. 11.1).

<Off>

Function deactivated

<On>

Function activated

Procedure when function is enabled:

Prior to starting the measurement,
the instrument starts heating up on
the set drying temperature.

If necessary close the sample
chamber, preheating will start.

You may stop preheating during the
preheating period by tapping
and after weighing the sample
measurement can be started.



For standard applications preheating normally is not required.



Advertising