Put on head covers (C).

+

Pay attention to the correct orientation of the

head covers (see figure below).

+

Make sure that the diaphragms stay centered

with respect to the bores so that they will be-

come clamped uniformly between housing

(I) and head covers (C).

Put the valves (C) in place. See figure below for the correct position of

the valves:

Inlet side of pump head: The valve tongue points at the kidney-shaped

orifice in the valve seat (1).

Outlet side of pump head: The valve is oriented the opposite direc-

tion as the valve at the inlet side. Round orifice under the valve tongue

(2).

outlet

side (2)

inlet

side (1)

Scheme of pump heads with head covers and valves