Connecting components at the controller:

At the rear side of the controller are connections

for e.g., an optional coolant valve and/or an ex-

ternal pressure transducer and/or an external

venting valve, as well as a venting connection

and the serial interface RS 232C.

- Pull the controller out of the housing. Be careful not to

disconnect the measuring connection (PTFE tube)!

- Insert the cable of the component to be connected un-

derneath the housing and plug into the controller. Do

not apply off-axis forces when assembling or removing

plug connections! Observe correct orientation of the

plug.

- Insert controller into the housing.

Separator at inlet and exhaust waste vapor condens-

er:

Assembling the hose nozzle with union nut:

Take the hose nozzle with attached compression fer-

rule and union nut out of the catchpot and put onto inlet

connection.

Tighten the union nut by hand until you can feel the

stop. Then tighten an additional 1/4 rotation with an

open-ended wrench (size 17mm) for final installation.

Catchpots:

The catchpot at the inlet protects against drop-

lets and particles from entering the pump.
+

Enhances lifetimes of diaphragms and

valves.

+

Improves vacuum performance in applica-

tions with condensable vapors.

Both catchpots are coated with a protective lay-

er to protect against shattering in case of break-

age or implosion.

catchpot at

the inlet (10)

catchpot at the outlet (10)