Attention: Coolant may overflow unexpectedly, e.g., in

case of large amounts of gas.

+

Do not overfill the condenser. Lowest permissible cool-

ant temperature: -112°F (-80°C).

+

Maximum operating pressure: 16 psi (1.1 bar) absolute.

• Ensure that no parts of your clothing, hair or fingers can

be caught or drawn in at the inlet of the pump. Never

insert fingers or drop any other object into the inlet or

outlet.

• Pumping at high inlet pressure may lead to overpres-

sure at the gas ballast valve. Pumped gases or con-

densate might be expelled if the valve is open. If an

inert gas supply is connected to the gas ballast, ensure

that its inlet pipeline is not contaminated.

• You must take suitable precautions to prevent any dan-

gerous situation from arising if the controller starts the

pump, switches a coolant valve or opens a venting

valve.

Attention: If the controller is set to Autostart, the pro-

cess will start immediately after a power failure with-

out pressing any further key. It is your responsibility to

ensure that automatic start-up of the system will not

lead to any dangerous condition. Provide appropriate

safety measures. Check

prior to starting the process

whether the option ”Autostart” (menu: configuration) is

enabled.

• Pay attention to the safety symbol ”hot surfaces” on

the equipment. Hot parts may cause burns if touched.

Adopt suitable measures to prevent any danger arising

from hot surfaces or electric sparks. Ensure that hot

surfaces of the pump do not cause burns. Provide a

suitable contact guard if necessary.

• Ensure that the

coolant outlet pipe at the waste vapor

condenser is always free and that it cannot get blocked.