Cleaning the pressure transducer, Readjustment of the controller cvc 3000 – VACUUBRAND PC 3001 VARIO emission condenser Peltronic User Manual

Page 83

Advertising
background image

page 83 of 120

Attention: Never use a pointed or sharp-edged tool

to clean the pressure transducer.

Never touch the ceramic diaphragm of the pressure

transducer with hard objects.

Fill the measurement chamber with a solvent (e.g.,

benzene) and allow sufficient cleaning time. Observe

all regulations concerning usage and disposal of sol-

vents!

Drain the solvent and dispose of in accordance with

regulations. Repeat cleaning if necessary.

Rinse the measurement chamber several times with al-

cohol in order to remove all solvent residues.

Allow the pressure transducer to dry.

Readjust the pressure transducer if necessary.

Readjustment of the controller CVC 3000

See section ”Readjustment of CVC 3000”, pg. 80.

Cleaning the pressure transducer