Do not pump:

-

unstable substances

- substances which react explosively under

impact

(mechanical stress) without air

- substances which react explosively when being ex-

posed to

elevated temperatures without air,

- substances subject to

auto-ignition,

- substances which are

inflammable without air

-

explosive substances.

The pumps are

not approved for operation below

ground.

Do not operate the pump below ground.

+

The pumps are

not suitable for pumping dust.

Do not pump dust.

Do not pump substances which may form deposits in-

side the pump. The pumps are not suitable for pumping

substances which may form deposits inside the pump.

Deposits and condensate in the pump may lead to in-

creased temperatures even to the point of exceeding

the maximum permitted temperatures.

• Check the inlet and outlet of the pump, if there is a dan-

ger of forming

deposits inside the pump, e.g., in the

pump chambers (the pump chamber is the part between

diaphragm and head cover. See section ”Replacing dia-

phragms and valves”, pg. 99). Inspect the pump cham-

bers regularly and clean if necessary.

Consider interactions and chemical reactions of

the pumped media. Ensure that the materials of the

pump’s wetted parts are compatible with the pumped

substances, see section ”Technical data”, pg. 27.

When changing the substances pumped, we recom-

mend purging the pump with air or inert gas prior to

changing the pumped media. Purging the pump will

pump out residues and it will reduce the possibility of

reactions of the pumped substances with each other

and with the pump’s materials.