Checking the ultimate vacuum – VACUUBRAND PC 3001 VARIO emission condenser Peltronic User Manual

Page 111

Advertising
background image

page 111 of 120

PC 3001 VARIO EK Peltronic

Assemble the emission condenser Peltronic.

Slip the hose from the outlet of the pump

onto the hose connector of the emission con-

denser Peltronic.

Attach the hose connection with hose clip.

Close hose clip with flat pliers.

Attach the mounting plate of the emission

condenser Peltronic with 2 screws at the

housing cover of the pumping unit; use a

Phillips screw driver size 2. Pay attention to

the cage nuts in the groove of the housing

cover.

Slip connecting hose onto hose connection

of fitting below the housing cover.

Close hose clip with flat pliers.

Assemble catchpots (10) with joint clips.

Checking the ultimate vacuum

After any intervention at the equipment (e.g., repair / maintenance)

the ultimate vacuum of the pump has to be checked. Only if the pump

achieves its specified ultimate vacuum, the pump’s leak rate is low

enough to ensure that no explosive atmospheres will occur in the inte-

rior of the equipment.

After any intervention at the vacuum sensor the leak rate of the equip-

ment has to be checked.