Cleaning and replacing components

Replacing the overpressure safety relief device:

Loosen the union nut at the inlet of the exhaust vapor condenser.

Pull the PTFE hose out of the inlet of the exhaust vapor condenser.

If necessary, pull off the old pressure relief valve and install the new

one. Check for correct position of the PTFE-film under the pressure

relief valve.

Feed the PTFE hose into the inlet of the exhaust vapor condenser.

Fasten union nut.

Overpressure

safety relief device ............................ 638821

(at exhaust waste vapor condenser)

Catchpot

0.52 qt (500 ml), coated................................................................

638497

O-ring 28 x 2.5 ........................................................................... 3120553

(FPM) at the spherical ground joint of the catchpot at the inlet

O-ring 28 x 2.5 ............................................................................. 638304

FEP bonded for increased chemical resistance

If the pump does not achieve the ultimate vacuum:

- Whenever the diaphragms and valves have been replaced, a break-in

period of several hours is required before the pump achieves its ulti-

mate vacuum.

- In case of an unusual noise, switch off pump immediately and check

clamping disc positions.

If the specified ultimate vacuum is not achieved, and if this does not

change after the break-in period:

Check hose connectors at pump heads for leaks. If necessary recheck

valve seats and pump chambers.