Important notes concerning the operation of the, Exhaust waste vapor condenser, Notice – VACUUBRAND PC 3001 VARIO emission condenser Peltronic User Manual

Page 49

Advertising
background image

page 49 of 120

To reduce condensation in the pump, do not pump vapor

before the pump has reached its operating temperature.

Open the gas ballast valve when pumping condensable

vapors. Turn gas ballast cap to open valve.

For

condensable vapors (water vapor, sol-

vents, etc.):

- The gas ballast valve is open if the arrow on

the gas ballast cap is pointing towards the

pump.

- With gas ballast valve open, the ultimate

vacuum will be reduced.

- Use inert gas for gas ballast to avoid the formation of

explosive mixtures.

Attention: maximum supply pres-

sure of inert gas: 17.5 psi (1.2 bar) absolute.

- Close the gas ballast valve by turning the cap 180°.

In case of low boiling solvents (when the formation of con-

densate is unlikely), the use of gas ballast might be unnec-

essary. Operating the pump without gas ballast increases

the solvent recovery rate at the exhaust waste vapor con-

denser.

Important notes concerning the operation of

the exhaust waste vapor condenser

Connect the exhaust to a suitable treatment system to

prevent the discharge of dangerous gases and vapors

to the surrounding atmosphere.

+

Never block the gas outlet ((6) hose nozzle for tubing

I.D. 3/8” (10 mm)). The exhaust hose must always be

unobstructed and without back-pressure to enable an

unhindered discharge of gases.

+

Check the overpressure safety relief device (17) at the

exhaust waste vapor condenser (9) regularly; replace if

necessary. Check especially for deterioration, coales-

cence and cracks.

NOTICE

gas

ballast