Important notes regarding the use of gas ballast – VACUUBRAND PC 3001 VARIO emission condenser Peltronic User Manual

Page 48

Advertising
background image

page 48 of 120

the silencer regularly and replace if necessary.

A temperature sensor at the circuit board protects the mo-

tor: Current limitation in the event the temperature at the

circuit board raises above 158°F (70°C). At temperatures

above 185°F (85°C) the pump is switched off. In the event

of a motor blockage (after 10 start-up attempts) the pump

is switched off.

Note: Only manual reset is possible. Disconnect the pump

from the power source. Identify and eliminate the cause of

failure.

Check ventilation slots regularly for dust/dirt. Clean if nec-

essary. Avoid overheating (e.g., due to hot process gas-

es).

A warm up period (approximately 15 min.) is required to

ensure that the rated ultimate vacuum and pumping speed

are attained.

Important notes regarding the use of gas bal-

last

Gas ballast is a continuous purge to keep the pump’s in-

terior as clean as possible and to reduce the possibility of

condensation inside the pump.

Air and pumped media might react inside the pump or

at the outlet of the pump and form hazardous or explo-

sive mixtures, when you use air rather than inert gas

for the gas ballast. This constitutes a risk of significant

damage to equipment and/or facilities, a risk of person-

al injury or even loss of life.

+

Make sure that air/gas intake through the gas ballast

valve can never lead to hazardous, explosive or other-

wise dangerous mixtures. If in doubt, use inert gas.