if installing the pump in a cabinet or a housing. Make

sure ventilation is adequate to maintain recommended

operating temperature. Install an external automatic

ventilation system if necessary. If pumping hot process

gases, make sure that the maximum permitted gas in-

let temperature is not exceeded. The maximum permit-

ted gas inlet temperature depends on several param-

eters like inlet pressure and ambient temperature (see

”Technical data”, pg. 27).

+

Do not aspirate particles and dust.

+

Do not pump liquids.

• Ensure that the pump, and any accessories in the flow

path are chemically resistant to the pumped substanc-

es prior to operation.

Use the equipment

only as intended, that is, for genera-

tion, measurement, and control of vacuum in vessels de-

signed for that purpose. Any other use will automatically

invalidate all warranty and liability claims. Remain aware

of safety and risks.

Setting up and installing the equipment

Equipment must be connected only to a

suitable elec-

trical supply and a suitable ground point. As such,

the plug must be plugged into an outlet that is properly

grounded. Failure to connect the motor to ground may

result in deadly electrical shock.

The supply cable may be fitted with a molded Euro-

pean IEC plug or a plug suitable for your local electrical

supply. The cable contains wires color coded as fol-

lows: green or green and yellow: ground; blue or white:

neutral; brown or black: hot.

+

Due to the high compression ratio, the pump may gen-

erate overpressure at the outlet. Check pressure com-

patibility with system components (e.g., exhaust pipe-

NOTICE