• Ensure that the

coolant outlet pipe is always free and

that it cannot get blocked. If installing an optional cool-

ant valve, it must always be in the supply line of the

exhaust waste vapor condenser.

Keep a minimum distance of 8 in (20 cm) between the

cooling fan and surrounding items (e.g., housing, walls,

etc.). Clean ventilation slots if necessary to avoid a reduc-

tion of ventilation.

Do not place the pumping unit on soft surfaces (e.g., rub-

ber foam) during operation. This may reduce or block the

electronics’ air supply. Do not cover the pumping unit.

Use only hoses at the inlet and outlet of the pump with

an inner diameter at least as large as the diameter of the

pump’s tubing (to avoid overpressure at the outlet, and

reduction of pumping speed at the inlet).

Allow the equipment to equilibrate to ambient temperature

if you bring it from cold environment into a room prior to

operation. Notice if there is water condensation on cold

surfaces.

Secure coolant hoses at the hose nozzles (e.g., with hose

clamp) to prevent their accidental slipping.

Comply with all

applicable and relevant safety require-

ments (regulations and guidelines). Implement the re-

quired actions and adopt suitable safety measures.

Ambient conditions

Do not reach for this product if it has fallen into liquid.

There is a risk of deadly electrical shock. Unplug the

system immediately.

+

Do not use this product in an area where it can fall or

be pulled into water or other liquids.