Notes regarding the (dry) ice condenser – VACUUBRAND PC 3001 VARIO emission condenser Peltronic User Manual

Page 43

Advertising
background image

page 43 of 120

Note: Install the hoses of the cooling system in a way

to avoid the flow / dripping of condensed water onto the

pumping unit (especially cables and electronic parts,

see also IP degree of protection, ”Technical data”, pg.

27.

• Ensure that the

coolant outlet tubing is always unob-

structed and that it cannot get blocked.

• Maximum permissible coolant pressure at the exhaust

waste vapor condenser: 87 psi (6 bar) absolute. Outlet

flow must always be unhindered.

• Comply with the maximum permissible coolant pres-

sures of additional components in the coolant circuit

(e.g., coolant valve).

• Avoid overpressure in the coolant circuit (e.g., caused

by blocked or squeezed coolant hoses).

• Only install the optional coolant valve in the supply line

of the exhaust waste vapor condenser.

Notes regarding the (dry) ice condenser

(PC 3001 TE VARIO)

The

exhaust waste vapor condenser enables

an efficient condensation of the pumped vapors

at the outlet.
+

Close to 100% solvent recovery.

+

The isolation cover protects against glass

splinters in case of breaking, acts as thermal

insulation to avoid the condensation of hu-

midity and is intended to absorb shocks.

Adopt suitable measures to prevent the formation of

explosive or flammable mixtures, use inert gas for vent-

ing if necessary.

Prevent the discharge of dangerous gases and vapors

to the surrounding atmosphere. If appropriate, connect

the exhaust line to a suitable treatment system.

Adopt suitable measures to prevent the release of dan-

gerous, toxic, explosive, corrosive, noxious or polluting

fluids, vapors and gases when disposing of conden-

sates.

outlet (gas!)

hose nozzle