When assembling, ensure

vacuum-tightness. After as-

sembly, check the whole system for leaks.

Secure hose connections at the pump appropriately, e.g.,

with hose clamps, to protect against accidental detach-

ment.

To reduce pump noise emanating from the pump exhaust

port, connect an exhaust hose or use a silencer (see ”Ac-

cessories”, pg. 94).

The VACUUBRAND controller CVC 3000 can only be oper-

ated with components compatible with the VACUUBRAND

VACUU•BUS system, (see ”Accessories”, pg. 94).

The vacuum controller CVC 3000 controls VACUUBRAND

diaphragm pumps NT VARIO and VARIO-B, pumping

units PC 30xx VARIO, and optional coolant and venting

valves. Connected components (e.g., venting valve, level

sensor, external pressure transducer 3000 series) are au-

tomatically identified and configurated. Identical compo-

nents must be configurated beforehand; information upon

request.

Do not use more than one controller within the same

VACUU•BUS system.

Several controllers in the same VACUU•BUS sys-

tem will interfere with each other and result in error

messages of the connected components (pumps,

valves).

CVC 3000

The CVC 3000 is equipped with an internal capacitive

pressure transducer with ceramic diaphragm. It measures

the actual pressure independently of the gas type, and

with reference to the vacuum, i.e., absolute.

- Condensate and deposits will affect the mea-

surement results. Clean the pressure trans-

ducer, if necessary. See section ”Cleaning

the pressure transducer”, pg. 83.

measurement

connection

at CVC 3000