• Comply with

maximum permissible pressures at the

pressure transducer. See section “Technical data”, pg.

26.

• Avoid overpressure of more than 17.5 psi absolute (1.2

bar absolute) in the event that inert gas is connected to

the pump, or to a venting valve.

Note: Flexible elements will shrink when evacuated.

• Connect hoses gas tight at inlet and outlet of the pump.

• Connect hoses gas tight at the pressure transducer.

• Ensure that no foreign objects can be drawn into the

pump.

• Check the power source and the pump’s rating plate

to be sure that the power source and the equipment

match in voltage, phase, and frequency.

Make sure ventilation is adequate to maintain recommend-

ed operating temperature. Keep a minimum distance of

2 in (5 cm) between the cooling fan and surrounding items

(e.g., housing, walls, etc.), or else install an external au-

tomatic ventilation system. Do not place the pump on soft

surfaces (e.g., rubber foam) during operation. This may

reduce the air circulation in the pump housing and affect

the fans’ cooling effect. Check fans regularly for dust/dirt.

Clean fan guard grills if necessary to avoid a reduction of

ventilation.

Position the pressure transducer and its vacuum line in

such a way that condensate cannot flow towards the pres-

sure transducer.

Use only hoses at the inlet and outlet of the pump with

an inner diameter at least as large as the diameter of the

pump’s tubing (to avoid overpressure at the outlet, and

reduction of pumping speed at the inlet).